Continuµm User Guide 135
The microscope’s optical design is symmetrical with respect to the
sample and aperture. This lets you use a single aperture for masking
both before and after the sample. Also, the pre-and post-sample
aperture size and orientation are identical. The pre-sample aperture
position is fixed in relation to the objective in both transmission
mode and reflection mode. In transmission mode the aperture image
is in focus when the condenser image is in focus; in reflection mode
the aperture image is in focus when the sample is in focus.
In transmission mode, you adjust the post-sample aperture image by
adjusting the condenser: centering and focusing the condenser also
centers and focuses the post-sample aperture image. See “Condenser
focus and alignment” in the “Preparing the Microscope” chapter for
instructions.
Adjusting the
aperture size,
shape and orientation
For systems with an optional automated Reflex aperture, the size,
shape, and orientation of the Reflex aperture can be set by:
•
Loading an experiment file that has an aperture setting included
in the saved parameters.
•
Using the mouse to point to a side or unmarked corner of the
aperture box in the video pane and dragging it to the size, shape,
and orientation you wish to use.
•
Setting the size and rotation in the Aperture Dimensions dialog.
For all systems, the aperture can be set by physically changing the
setting of the Reflex aperture controls on the right-side of the
microscope.
Notice
The aperture automation software cannot track changes made using
the knobs on the right-side of the microscope. To synchronize the
physical aperture with the automated aperture system choose System
Configuration from the Atlµs menu and then press the Aperture
button in the Initialization field. If you have used the manual
controls and do not synchronize the aperture with the software
before loading an experiment with automated aperture settings or
attempting to use the software to adjust the aperture, the size and
shape of the aperture may not adjust correctly.
Содержание Nicolet Continuum
Страница 1: ......
Страница 9: ...Continuµm User Guide 5 Front panel ...
Страница 10: ...Thermo Scientific 6 Control panel on left side ...
Страница 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Страница 48: ...Thermo Scientific 44 ...
Страница 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Страница 80: ...Thermo Scientific 76 17 When data collection finishes use the viewer pane to isolate peaks of interest ...
Страница 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Страница 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Страница 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Страница 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Страница 126: ...Thermo Scientific 122 ...
Страница 151: ...Continuµm User Guide 147 5 Increase the reflection illumination so that it is brighter than normal viewing intensity ...
Страница 154: ...Thermo Scientific 150 ...
Страница 176: ...Thermo Scientific 172 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...
Страница 183: ...Continuµm User Guide 179 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...