Thermo Scientific
108
18. When data collection is finished, follow the instructions that
appear on the screen to display the ratioed sample spectrum
in a spectral window.
This spectrum shows the results of your analysis. (If Collect To
A New Window is turned on in the Collect options, the spectrum
is displayed in a new spectral window automatically.)
If you need help, find “sample spectrum” in the OMNIC Help
system Index and go to “How to collect a sample spectrum.”
Setting up the
external quantitative
Contact Alert System
The external quantitative Contact Alert System has its own
indicators, an adjustable audible alarm, and an analog meter for
measuring the contact pressure. While you are using the external
Contact Alert system, the Contact and Stop indicators on the front
panel of the microscope do not light; they are replaced by the visual
and audible alarms available through the controller.
Notice
The contact alarm lines for this system are not monitored by
software. Always disable Auto ATR while using the external
qualitative Contact Alert System. Damage resulting from stage
crashes is not covered by warranty.
See
the Contact Alert User’s Manual for detailed procedures for
using the quantitative Contact Alert system.
Use the following procedure to set up the external, quantitative
Contact Alert system. Use this option when you need to measure
samples with the best possible reproducibility of contact pressure;
for example, for quantitative analyses or research studies.
Caution
Avoid shock hazard. Before connecting the quantitative Contact
Alert System, always turn off the microscope power.
Содержание Nicolet Continuum
Страница 1: ......
Страница 9: ...Continuµm User Guide 5 Front panel ...
Страница 10: ...Thermo Scientific 6 Control panel on left side ...
Страница 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Страница 48: ...Thermo Scientific 44 ...
Страница 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Страница 80: ...Thermo Scientific 76 17 When data collection finishes use the viewer pane to isolate peaks of interest ...
Страница 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Страница 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Страница 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Страница 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Страница 126: ...Thermo Scientific 122 ...
Страница 151: ...Continuµm User Guide 147 5 Increase the reflection illumination so that it is brighter than normal viewing intensity ...
Страница 154: ...Thermo Scientific 150 ...
Страница 176: ...Thermo Scientific 172 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...
Страница 183: ...Continuµm User Guide 179 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...