Continuµm User Guide 155
Symptom
Possible Cause
Solution
Video pane of the OMNIC
Atlµs or µView window is
dark grey or black,
continued…
Field of view and/or contrast
controls are improperly
adjusted.
Adjust the field and aperture iris settings.
See
“Adjusting field of view and contrast” in the
“More about the hardware” chapter.
Microscope and/or illuminators
are turned off.
Verify that the microscope and illuminator
power is ON.
See
“Turning on the
microscope power” in “Preparing the
Microscope” chapter.
Verify that the power cords are connected to
a working AC source (wall outlet or power
strip) and firmly seated at both ends.
Illuminators are improperly
adjusted.
Adjust the illuminator intensity.
See
“Preparing the optics” in “Preparing the
Microscope” chapter.
No video signal is reaching the
computer.
Verify that the microscope and illuminator
power is ON.
See
“Turning on the
microscope power” in “Preparing the
Microscope” chapter.
Verify that the power cords are connected to
a working AC source (wall outlet or power
strip) and firmly seated at both ends. Replace
any damaged cables.
Verify that the video cable (composite or S
Video) is connected between the camera and
the microscope (analog) or the USB cable is
connected from the microscope to the
computer (digital)..
Replace any damaged cables.
continued…
Содержание Nicolet Continuum
Страница 1: ......
Страница 9: ...Continuµm User Guide 5 Front panel ...
Страница 10: ...Thermo Scientific 6 Control panel on left side ...
Страница 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Страница 48: ...Thermo Scientific 44 ...
Страница 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Страница 80: ...Thermo Scientific 76 17 When data collection finishes use the viewer pane to isolate peaks of interest ...
Страница 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Страница 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Страница 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Страница 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Страница 126: ...Thermo Scientific 122 ...
Страница 151: ...Continuµm User Guide 147 5 Increase the reflection illumination so that it is brighter than normal viewing intensity ...
Страница 154: ...Thermo Scientific 150 ...
Страница 176: ...Thermo Scientific 172 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...
Страница 183: ...Continuµm User Guide 179 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...