Thermo Scientific
120
11. Set the experiment parameters.
Use Experiment Setup in the collect menu to test the parameters.
(If you have set up and saved an experiment for GAO data
collection, you can set the parameters in one step by selecting or
opening the experiment.) Set Background Handling on the
Collect tab to Collect Background After Every Sample.
Note
Grazing angle measurements may require more than 1000 scans to
achieve an adequate signal-to-noise ratio, especially when band
intensities are weak.
If you need help setting parameters, find “experiment” in the
OMNIC Help system Index and go to “Using Experiment Setup.”
12. Choose collect Sample from the Collect menu to collect the
sample spectrum.
Follow the instructions that appear on the screen.
13. When you are prompted to collect the background spectrum,
position under the GAO a clean area of the reflective
substrate on which the sample is installed.
Do not adjust the Reflex aperture or the GAO.
14. While looking through the viewer (or at the video image),
finely focus the image of the substrate surface.
Содержание Nicolet Continuum
Страница 1: ......
Страница 9: ...Continuµm User Guide 5 Front panel ...
Страница 10: ...Thermo Scientific 6 Control panel on left side ...
Страница 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Страница 48: ...Thermo Scientific 44 ...
Страница 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Страница 80: ...Thermo Scientific 76 17 When data collection finishes use the viewer pane to isolate peaks of interest ...
Страница 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Страница 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Страница 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Страница 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Страница 126: ...Thermo Scientific 122 ...
Страница 151: ...Continuµm User Guide 147 5 Increase the reflection illumination so that it is brighter than normal viewing intensity ...
Страница 154: ...Thermo Scientific 150 ...
Страница 176: ...Thermo Scientific 172 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...
Страница 183: ...Continuµm User Guide 179 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...