46
The aperture automation software cannot track manual changes to
the aperture (changes made using the knobs on the right side of the
microscope). Any time you make manual changes to the aperture,
you must re-synchronize the hardware and software controls.
Notice
If you do not synchronize the aperture with the software before
loading an experiment with aperture settings or attempting to use the
software to adjust the aperture, the size and shape of the aperture
may be unpredictable.
To synchronize the physical aperture with the automated aperture
system:
1. Open OMNIC and select Transmission discrete point –
standard sample from the Experiment drop-down list box.
2. Choose Show Atlµs Window from the Atlµs menu to open the
Atlµs window.
When the Atlµs window opens, you will see navigation pane
(left side of screen) that shows the current stage location and a
video pane (right side of screen).
The colored box in the video pane represents the virtual aperture.
Содержание Nicolet Continuum
Страница 1: ......
Страница 9: ...Continuµm User Guide 5 Front panel ...
Страница 10: ...Thermo Scientific 6 Control panel on left side ...
Страница 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Страница 48: ...Thermo Scientific 44 ...
Страница 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Страница 80: ...Thermo Scientific 76 17 When data collection finishes use the viewer pane to isolate peaks of interest ...
Страница 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Страница 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Страница 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Страница 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Страница 126: ...Thermo Scientific 122 ...
Страница 151: ...Continuµm User Guide 147 5 Increase the reflection illumination so that it is brighter than normal viewing intensity ...
Страница 154: ...Thermo Scientific 150 ...
Страница 176: ...Thermo Scientific 172 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...
Страница 183: ...Continuµm User Guide 179 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...