Continuµm User Guide
11
Cooling the detector
The microscope is equipped with one or two single-element
detectors. Most of these detectors must be cooled with liquid
nitrogen before use.
Caution
Liquid nitrogen is extremely cold and can cause injury. A warm
laboratory dewar, funnel, and detector may cause the liquid nitrogen
to boil rapidly and spatter. When filling the detector dewar, be
careful not to contact the liquid nitrogen with your skin.
Wear protective gloves and splash-proof goggles and follow
precautions described in the material safety data sheets provided by
your liquid nitrogen vendor.
Also note that liquid nitrogen can be hazardous and create an oxygen
deficient atmosphere if handled in an insufficiently ventilated
room.
Detectors that require cooling are encased in a stainless steel dewar.
Once cooled, the detector will remain cold for about 13 hours. You
can add liquid nitrogen as needed at any time except while you are
collecting spectra.
To cool a detector:
1. Fill a laboratory dewar with approximately 750 ml (26 fluid
ounces) of liquid nitrogen.
Less liquid nitrogen may be needed if the detector was cooled
recently.
Содержание Nicolet Continuum
Страница 1: ......
Страница 9: ...Continuµm User Guide 5 Front panel ...
Страница 10: ...Thermo Scientific 6 Control panel on left side ...
Страница 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Страница 48: ...Thermo Scientific 44 ...
Страница 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Страница 80: ...Thermo Scientific 76 17 When data collection finishes use the viewer pane to isolate peaks of interest ...
Страница 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Страница 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Страница 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Страница 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Страница 126: ...Thermo Scientific 122 ...
Страница 151: ...Continuµm User Guide 147 5 Increase the reflection illumination so that it is brighter than normal viewing intensity ...
Страница 154: ...Thermo Scientific 150 ...
Страница 176: ...Thermo Scientific 172 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...
Страница 183: ...Continuµm User Guide 179 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...