Continuµm User Guide
49
Analyzing Samples
This chapter explains how to use a Nicolet Continuµm microscope
and OMNIC Atlµs to collect infrared spectra of samples in reflection
and transmission modes. Use the standard experiments and the
samples that came with the microscope to perform these experiments
and familiarize yourself with the microscope controls and software.
Performing
a reflection
experiment
A standard reflection experiment sample was shipped in the
accessory box. This sample can be used to perform your first discrete
point reflection experiment and whenever you wish to confirm that
your microscope is operating properly.
1. Locate the Soda Can sample slide and install the slide in the
universal holder and then install the holder on the stage.
The slide was provided with your microscope in the accessory
box. It includes two open beam positions and an ink on brushed
aluminum sample.
2. Set the refractive index compensation for the objective.
Since this sample is a thin layer of ink on a reflective surface, set
the ring to 0.
Содержание Nicolet Continuum
Страница 1: ......
Страница 9: ...Continuµm User Guide 5 Front panel ...
Страница 10: ...Thermo Scientific 6 Control panel on left side ...
Страница 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Страница 48: ...Thermo Scientific 44 ...
Страница 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Страница 80: ...Thermo Scientific 76 17 When data collection finishes use the viewer pane to isolate peaks of interest ...
Страница 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Страница 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Страница 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Страница 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Страница 126: ...Thermo Scientific 122 ...
Страница 151: ...Continuµm User Guide 147 5 Increase the reflection illumination so that it is brighter than normal viewing intensity ...
Страница 154: ...Thermo Scientific 150 ...
Страница 176: ...Thermo Scientific 172 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...
Страница 183: ...Continuµm User Guide 179 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...