Continuµm User Guide 159
Symptom
Possible Cause
Solution
No image appears in the video
pane or the viewer,
continued...
One or more illuminator bulbs
are burned out.
Verify that bulbs are still functioning by
adjusting illumination to maximum intensity
and looking for light inside bulb access
covers.
See
“Replacing an illuminator bulb”
in the Service and Maintenance” Help book,
available through “Microscope Help topics”
in the OMNIC Help menu for step-by-step
instructions.
Microscope power is turned off
or power cord is disconnected.
Make sure power cord is connected and
power switch is turned on (“I” position).
See
“Turning on the microscope power” in
“Preparing your Microscope” chapter.
A fuse is blown.
Check fuses and replace any fuse that is
blown.
See
“Replacing a fuse” in
“Microscope Help topics” in the OMNIC
Help menu.
Video signal is not reaching the
computer.
Verify that the video cable (composite or S
Video) is connected between the camera and
the microscope and between the microscope
and computer (analog) or the USB cable is
connected from the camera to the
microscope and from the microscope to the
computer (digital). Be sure it is firmly seated
at all ends.
Video cables have been
damaged
Inspect and replace any faulty cables.
The microscope, camera,
computer, and/or video capture
card are malfunctioning.
Contact technical support.
continued…
Содержание Nicolet Continuum
Страница 1: ......
Страница 9: ...Continuµm User Guide 5 Front panel ...
Страница 10: ...Thermo Scientific 6 Control panel on left side ...
Страница 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Страница 48: ...Thermo Scientific 44 ...
Страница 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Страница 80: ...Thermo Scientific 76 17 When data collection finishes use the viewer pane to isolate peaks of interest ...
Страница 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Страница 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Страница 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Страница 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Страница 126: ...Thermo Scientific 122 ...
Страница 151: ...Continuµm User Guide 147 5 Increase the reflection illumination so that it is brighter than normal viewing intensity ...
Страница 154: ...Thermo Scientific 150 ...
Страница 176: ...Thermo Scientific 172 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...
Страница 183: ...Continuµm User Guide 179 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...