Continuµm User Guide
53
6. Use the joystick or Move Stage tools to move the sample into
the field of view and then use the Z-axis or stage focus knob
sharply focus on the ink pattern.
You should now be able to see a clear image of the ink in the
video pane.
Note
A yellow circle in the Bench Status field usually indicates that the
detector has not been cooled. If you already added liquid nitrogen to
the detector dewar, simply wait for temperature to stabilize. If you
have not yet cooled the detector, do so before attempting to collect
data.
You may also see a yellow circle in the Bench Status field if the
quality checks, available through the Quality tab in Experiment
Setup have not been disabled. The default quality check settings
were designed for experiments conducted in the spectrometer sample
compartment. They are not useful for sampling with the
microscope.
7. Set the irises.
Fully open the reflection field iris.
Содержание Nicolet Continuum
Страница 1: ......
Страница 9: ...Continuµm User Guide 5 Front panel ...
Страница 10: ...Thermo Scientific 6 Control panel on left side ...
Страница 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Страница 48: ...Thermo Scientific 44 ...
Страница 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Страница 80: ...Thermo Scientific 76 17 When data collection finishes use the viewer pane to isolate peaks of interest ...
Страница 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Страница 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Страница 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Страница 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Страница 126: ...Thermo Scientific 122 ...
Страница 151: ...Continuµm User Guide 147 5 Increase the reflection illumination so that it is brighter than normal viewing intensity ...
Страница 154: ...Thermo Scientific 150 ...
Страница 176: ...Thermo Scientific 172 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...
Страница 183: ...Continuµm User Guide 179 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...