Thermo Scientific
176
12. When you are prompted to collect the background spectrum,
choose OK.
13. When data collection finishes, follow the instructions that
appear on the screen to display the ratioed sample spectrum
in a spectral window.
If Collect To a New Window is turned on in the Collect options,
the spectrum is displayed in a new spectral window
automatically.
14. Choose Noise from the Analyze menu.
The Noise dialog box appears. The peak-to-peak noise should be
less than 0.09 for MCT-A detectors using a 15X objective. For
other detectors, contact technical support to discuss areas of
concern. (Older spectrometers that do not meet basic mid-IR
performance standards may have a higher noise level.) If the
noise exceeds this level, discuss the matter with your service
representative.
Find
“noise” in the OMNIC Help system Index and go to
“Measuring noise,” if you need more information about the Noise
Command.
In transmission mode
Follow these steps to run the noise test in transmission mode:
1. Open the Atlµs window and initialize the Reflex aperture.
This gives an aperture size of 100 by 100 µm if you are using the
15X objective, or a size of 46.875 by 46.875 µm if you are using
the 32X objective.
Содержание Nicolet Continuum
Страница 1: ......
Страница 9: ...Continuµm User Guide 5 Front panel ...
Страница 10: ...Thermo Scientific 6 Control panel on left side ...
Страница 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Страница 48: ...Thermo Scientific 44 ...
Страница 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Страница 80: ...Thermo Scientific 76 17 When data collection finishes use the viewer pane to isolate peaks of interest ...
Страница 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Страница 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Страница 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Страница 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Страница 126: ...Thermo Scientific 122 ...
Страница 151: ...Continuµm User Guide 147 5 Increase the reflection illumination so that it is brighter than normal viewing intensity ...
Страница 154: ...Thermo Scientific 150 ...
Страница 176: ...Thermo Scientific 172 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...
Страница 183: ...Continuµm User Guide 179 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...