Function
SCPI Command
Description
Parameter and
Range
Query
Calibrate the
Current APG
Input
[:]CALibration:INPut:ANALog:CURRent
Calibrates the isolated and
non-isolated CAP
measurement commands.
<?|NR1>
Yes
Calibrate the
Voltage
controlled Non
Isolated
Voltage APG
feature
[:]CALibration[<channel>]:OUTPut
:ANALog[:VOLTage]
Calibrate the voltage analog
programming system.
? | <NR1> | <nothing
for interactive mode>
Yes
Calibrate the
Voltage
controlled
Isolated
Voltage APG
feature
[:]CALibration[<channel>]:OUTPut
:ANALog[:VOLTage]:ISOLated
Calibrates the voltage analog
programming system.
? | <NR1> | <nothing
for interactive mode>
Yes
Calibrate the
Voltage
controlled Non
Isolated
Current APG
feature
[:]CALibration[<channel>]:OUTPut
:ANALog:CURRent
Calibrate the current analog
programming system.
? | <NR1> | <nothing
for interactive mode>
Yes
Calibrate the
Voltage
controlled
Isolated
Current APG
feature
[:]CALibration[<channel>]:OUTPut
:ANALog:CURRent:ISOLated
Calibrates the current analog
programming system.
? | <NR1> | <nothing
for interactive mode>
Yes
Calibrate the
Resistive
controlled Non
Isolated
Voltage APG
feature
[:]CALibration[<channel>]:OUTPut
:ANALog[:VOLTage]:RESistive
Calibrate the Resistive
controlled voltage analog
programming system.
? | <NR1> | <nothing
for interactive mode>
Yes
Calibrate the
Resistive
controlled
Isolated
Voltage APG
feature
[:]CALibration[<channel>]:OUTPut
:ANALog[:VOLTage]:RESistive:ISOLated
Calibrates the Resistive
controlled voltage analog
programming system.
? | <NR1> | <nothing
for interactive mode>
Yes
Calibrate the
Resistive
controlled Non
Isolated
Current APG
feature
[:]CALibration[<channel>]:OUTPut
:ANALog:CURRent:RESistive
Calibrate the Resistive
controlled current analog
programming system.
? | <NR1> | <nothing
for interactive mode>
Yes