Series 2268 850 Watt DC Power Supplies Reference Manual
Remote Operation
2268S-901-01 Rev. A / January 2015
159
Repeating an Auto Sequence Program
An Auto Sequence Program can be configured to repeat any number of times or infinitely.
The default at power up is to have the repeat count default to once. The program will repeat if
the repeat count can be changed using the :
PROGram [<channel>] : REPeat
command. If
a program is running or pause the
: PROGram [<channel>] : REPeat
command will not
work.
Commands:
[:]PROGram[<channel>]:REPeat {?|<count>|INFinity}
Where:
<count> is the repeated count from 1 to 65 534.
INFinity is the character mnemonic for repeating without end.
For Example:
:PROG:STAR
*IDN?
*IDN?
:PROG:STOP:PROG:REP 2
:PROG:STAT RUN
The output after the last command might be as follows:
KEITHLEY INST.,2268-33-25, 1352A00499, 1.00,09/10/14
KEITHLEY INST.,2268-33-25, 1352A00499, 1.00,09/10/14
KEITHLEY INST.,2268-33-25, 1352A00499, 1.00,09/10/14
KEITHLEY INST.,2268-33-25, 1352A00499, 1.00,09/10/14