SCPI Commands for Output Control
Function
SCPI Command
Description
Parameter and
Range
Query
Set Current
Sharing Mode
[[:]]SOURce[<channel>]]:COMBine:CSHare
[: MODE]
Set the unit to operate in the
current share controller or
slave mode. If only a single
unit is being used it should be
set to controller.
?|CONTroller
|SLAVe
Yes
Set
(Immediate)
Current Set
point
[[:]SOURce[<channel>]]:CURRent[:LEVel]
[:IMMediate][:AMPLitude]
Set current set point
(immediate)
?|<Current Set
Point>|MAXimum|
MINimum
Yes
Set Constant
Power Set
point
[:]SOURce:POWer:LEVel
Set constant power set point
(immediate)
?|<Power Set Point> Yes
Set Constant
Power Mode
1
[:]SOURce:POWer:MODe
Set constant power mode
(Enable/Disable)
?|ON|OFF
Yes
Set
(Immediate)
Voltage Set
point
[[:]SOURce[<channel>]]:VOLTage[:LEVel]
[:IMMediate][:AMPLitude]
Change voltage set point
(immediate)
?|<Voltage Set
Point>|MAXimum|
MINimum
Yes
Set Over
Voltage
Protection
Level
[[:]SOURce[<channel>]]:VOLTage:PROTect
ion[:OVERvoltage][:LEVel]
Set the over voltage
protection level.
?|<OVP Set
Point>|MAXimum|
MINimum
Yes
Set Under
Voltage
Protection
Level
[[:]SOURce[<channel>]]:VOLTage:PROTect
ion:UNDer[:LEVel]
Set the under voltage
protection level.
?|<UVP Set
Point>|MAXimum|
MINimum
Yes