Calibration and Troubleshooting
Series 2268 850 Watt DC Power Supplies Reference Manual
204
2268S-901-01 Rev. A / January 2015
Non-isolated Voltage Programming of Current Calibration
Important:
For maximum accuracy at a specific APG level, follow the
calibration procedure again but use the intended operating voltage level in steps
3 and 4.
To calibrate the non-isolated voltage programming of current:
1.
Short the main output with a shunt. Connect a multimeter to measure the voltage
across the shunt.
2.
Connect the input voltage source to the non-isolated connector, current
programming (J1.10 - J1.7).
3.
Set the Voltage APG level to 4 V. The SCPI command is:
SYST:COMM:APR:LEV:CURR 4
4.
Set the input source voltage to 4.000 V.
5.
Set the input source as Voltage Non-Isolated for Current. The SCPI command is:
SYST:REM:SOUR:CURR AVOL
6.
Turn the main output On by pressing the OUTPUT ENABLE Main button on the
front panel or send the command:
:OUTP ON
The OUTPUT ENABLE Main button will illuminate.
7.
Enter the interactive calibration mode by sending the SCPI command:
CAL:OUT P:ANAL:CURR
8.
Press the terminal keys - or + to reduce or increase the main output current value.
Adjust the main output current to 100%.
9.
Press any key other than the + or - keys to exit from interactive calibration mode.
Содержание Keithley 2268 Series
Страница 102: ......