Series 2268 850 Watt DC Power Supplies Reference Manual
Remote Operation
2268S-901-01 Rev. A / January 2015
137
OPERation SHUTdown PROTection Status Register
Bit
Bit Weight Bit Name
Description
0
1
OVP
Over Voltage protection has tripped.
1
2
UVP
Under Voltage protection has tripped.
2
4
OCP
Over Current protection has tripped.
3
8
Not Used
Not Used
4
16
AC Off
AC failure protection has tripped.
5
32
OTP
Over Temperature protection has tripped.
6
64
Not Used
Not Used
7
128
Foldback
Foldback protection has tripped.
8
256
Fan
Fan failure protection has tripped.
Current Share Sub-Register
This register shows the state of the current share configuration, which can either be set
through the front panel Current Share Config menu, or through the SCPI command.
Important:
If current sharing is not being used, the unit must be
configured as a controller. Failure to do so will result in unpredictable
operation.
The SCPI command (s) for these instructions are:
[[:]SOURce]:COMBine:CSHare[:MODE]
OPERation CSHare Status Register
Bit
Bit Weight
Bit Name
Description
0
1
CONTroller
The power supply is configured to be a Current Share
CONTroller.
1
2
SLAVe
The power supply is configured to be a Current Share
Slave.