Series 2268 850 Watt DC Power Supplies Reference Manual
Calibration and Troubleshooting
2268S-901-01 Rev. A / January 2015
215
EMERGENCY SHUTDOWN
In an emergency, carry out these steps:
1.
Shut the power supply OFF immediately.
2.
Disconnect the mains supply.
3.
Disconnect the power supply from the load.
UNUSUAL OR ERRATIC OPERATION
If the power supply displays any unusual or erratic operation, follow these steps:
1.
Shut the power supply OFF immediately.
2.
Disconnect the power supply from the load and external programming.
3.
Test the power supply with no load, running the tests in “Step
4.
If the tests show that the power supply is functioning normally, check all load,
programming, and monitoring connections and circuits.
5.
Check the AC input for correct voltage and frequency.
If the problem is not solved after you have followed this procedure, or if the unit fails to
operate correctly upon retesting, call your service technician.
TROUBLESHOOTING FOR OPERATORS
Troubleshooting
Symptom
Check
Further Checks and Corrections
The power supply is not
operating.
is
displayed.
Flash memory check sum error.
Do a reset. See “Resetting the Power
Supply” on page 62.
Important:
The power supply will boot with default options and calibration
values.