Remote Operation
Series 2268 850 Watt DC Power Supplies Reference Manual
136
2268S-901-01 Rev. A / January 2015
OPERation Status Register
Bit
Bit Weight
Bit Name
Description
0
1
CALibrating
Indicates that the supply is in CALibration Mode.
1
2
SETTling
Not implemented
2
4
RANGing
Not implemented
3
8
SWEeping
Not implemented
4
16
MEASuring
Not implemented
5
32
Waiting for Arm
Not implemented
6
64
Waiting for Trigger
Not implemented
7
128
CORRecting
Not implemented
8
256
Not Used
Not Used
9
512
SHUTdown Summary
Reflects the summary of the SHUTdown
Sub-Register.
10
1024
Local Lockout
Front panel lockout
11
2048
Current Share Summary
Reflects the summary of the CSHare Sub-Register.
12
4096
Constant Voltage Mode
Reflects the output mode of the supply.
13
8192
Constant Current Mode
Reflects the output mode of the supply.
14
16384
Program Running
Indicates if an Autosequence program is currently
running.
The operation shutdown status register describes the cause of the power supply shutting
down the output. More than one bit may be active and multiple actions will be required to
restart the unit. The protection shutdown sub-register indicates which protection mechanisms
have caused the power supply to shut down.
OPERation SHUTdown Status Register
Bit
Bit Weight Bit Name
Description
0
1
PROTection
Reflects the summary of the PROTection
sub-register.
1
2
INTerlock
The power supply is shut down by INTerlock signal.
2
4
Not Used
Not Used
3
8
External Shutdown
The power supply is shut down by External Shutdown
signal.