Calibration and Troubleshooting
Series 2268 850 Watt DC Power Supplies Reference Manual
200
2268S-901-01 Rev. A / January 2015
NON-ISOLATED ANALOG PROGRAMMING CALIBRATION
Prior to this, the main output must be calibrated first.
Important:
In calibration commands, when + or - keys are expected, any other
key will exit from the calibration mode.
Non-isolated Voltage Monitoring Calibration
Important:
For maximum accuracy at a specific APG level, follow the
calibration procedure again but use the intended operating voltage level in steps
3 and 8.
To calibrate the non-isolated voltage monitoring:
1.
Disconnect the load and connect the voltmeter to the output.
2.
Connect the voltmeter to the non-isolated voltage monitoring output lines (J1.13 -
J1.7).
3.
Set the Voltage APG level to 4 V. Send the SCPI command:
SYST:COMM:APR:LEV 4
4.
Set the input source as Local. Send the SCPI command:
SYST:REM:SOUR LOC
5.
Set the main output voltage to 100% by sending the command:
SOUR:VOLT<ModelVolt>
or use the rotary adjust/Enter control.
6.
Turn the main output On by pressing the OUTPUT ENABLE Main button on the
front panel or send the command:
:OUTP ON
The OUTPUT ENABLE Main button will illuminate.
7.
Enter interactive calibration mode by sending the SCPI command:
CAL:MON
8.
Press on the terminal keys - or + to reduce or increase monitoring value. Adjust the
monitoring value displayed on the voltmeter that was set up in Step 1 to read as
close as possible to 4.000 V.
9.
Press any key other than the + or - keys to exit from interactive calibration mode.