Remote Operation
Series 2268 850 Watt DC Power Supplies Reference Manual
150
2268S-901-01 Rev. A / January 2015
RESET COMMAND
Important:
This command takes longer to execute than other commands. Be
sure to allow extra time (approximately two seconds) for the command to run
before sending another command..
The Reset command performs a device reset. The Reset command is the third level of reset in
a three level reset strategy, set out in IEEE 488.2 (see IEEE 488.2 standard, section 17.1.2).
The Reset command shall do the following:
1.
Set the device-specific functions to a known state that is independent of the
past-use history of the device.
2.
Force the device into the OCIS state, (see IEEE 488.2 standard, section 12.5.2).
3.
Force the device into the OQIS state, (see IEEE 488.2 standard, section 12.5.3).
The reset command explicitly shall NOT affect the following:
1.
The state of the IEEE 488.1 interface.
2.
The selected IEEE 488.1 address of the device.
3.
The Output Queue.
4.
The Standard Status Register Enable setting.
5.
The Standard Event Status Enable (SESR) setting.
6.
The Operation and Questionable SCPI status registers and their fan out registers.
7.
Calibration data that affects device specifications.
8.
The Protected User Data query response.
Commands:
*RST
[:]SYSTem[<channel>]:RESet