Local Operation
Series 2268 850 Watt DC Power Supplies Reference Manual
52
2268S-901-01 Rev. A / January 2015
USING OVER TEMPERATURE PROTECTION LOCK (OTP)
The OTP lock protects the power supply in the event of an over temperature alarm. This
alarm could be caused by ventilation restriction or overheating due to fan failure.
Two modes are available:
•
Auto recovery
(
OTP OFF) where the power supply turns on again after cooling down.
•
Latch mode (OTP ON) where you will need to reset the unit by turning the front panel
power switch to Off and then On again.
If the OTP is activated, the main output will turn off and
(over temperature
protection) will blink on the display. The OTP alarm is an alarm that will auto recover when
the temperature falls back into normal bounds.
Defining the OTP Mode
To define the OTP lock:
1.
Turn the 9-position mode control to the PRT position. PrO OUP is displayed.
2.
Turn the rotary Adjust/Enter control until
is displayed.
3.
Press the rotary Adjust/Enter control to display
.
4.
Turn the rotary Adjust/Enter control to change to On or Off.
5.
Press the rotary Adjust/Enter control.
The SCPI command (s) for these instructions are:
[:]SYSTem[<channel(s)>]:PROTection:LATCh
Resetting in Latch Mode
To reset the power supply if the OTP alarm is tripped and the OTP mode is set to On (latching
enabled), press and hold the rotary Adjust/Enter control for approximately 3 seconds.
The SCPI command (s) for these instructions are:
[:]OUTPut[<channel(s)>]:PROTection:CLEar
[:]SYSTem[<channel(s)>]:PROTection:LATCh