Local Operation
Series 2268 850 Watt DC Power Supplies Reference Manual
62
2268S-901-01 Rev. A / January 2015
RESETTING THE POWER SUPPLY
The reset is used to clear the parameters to the factory default values.
Soft Reset
The soft reset is used to set the parameters (see table below) to the
default values, but it does not reset the calibration constants
To perform a soft reset:
1.
Turn the power supply to Off then On.
When the unit is powering on,
is displayed on the output voltage and current
displays.
2.
Press the key combination of both OUTPUT ENABLE Main button and OUTPUT
ENABLE Aux button and hold continuously for 3 seconds in order to execute.
If the flash fail alarm triggers, it is also possible to reset the power supply. The front
panel should display
.
3.
Once a reset has been triggered, the output voltage display shows
for 1
second.
The model number will be displayed for 1 second and then return to normal mode.
After executing a reset, the power supply's settings are restored to the default settings but
retain the calibration data. Following lists the default settings.
Parameter
Setting
Address
1
Data rate (Kbps)
9.6
Communication mode
RS-232
Vout setting
0 V
Iout setting
0 A
Output
Off
AUX Out
Off
Trigger
Off
Ext. On/Off Polarity
Negative
Auto start mode
Off
AUX Auto start mode
Off
OVP
Maximum
UVP
0 V
Foldback trigger
None
Foldback delay
0.5 s
Current Share Mode
Controller
Alarm Output Latches
263 (0 x 107, all latches enabled)
Alarms Mask
2047 (0 x 7FF, all enabled)
Interlock
Disabled
Voltage Analog Programming
Off
Содержание Keithley 2268 Series