Calibration and Troubleshooting
Series 2268 850 Watt DC Power Supplies Reference Manual
208
2268S-901-01 Rev. A / January 2015
Isolated Voltage Programming of Voltage Calibration
Important:
For maximum accuracy at a specific APG level, repeat the
calibration procedure but use the intended operating voltage level in step 3.
To calibrate the isolated voltage programming of voltage:
1.
Disconnect the load and connect the voltmeter to the output.
2.
Connect the input voltage source to the Isolated connector (J3.3 and J3.6).
3.
Set the Voltage APG level to 4 V. The SCPI command is:
SYST:COMM:APR:LEV 4
4.
Set the input source voltage to 4.000 V.
5.
Set the input source as Voltage Isolated for Voltage. The SCPI command is:
SYST:REM:SOUR AVOL
6.
Turn the main output to On by pressing the OUTPUT ENABLE Main button on
the front panel or send the command:
:OUTP ON
The OUTPUT ENABLE Main button will illuminate.
7.
Enter the interactive calibration mode by sending the SCPI command:
CAL:OUTP:ANAL:ISOL
8.
Press on the terminal keys - or + to reduce or increase the main output voltage
value. Adjust the main output voltage to 100%.
9.
Press any key other than the + or - keys to exit from interactive calibration mode.