TE5251
User Manual
6-26
Input
Impedance:
50
Ω
Range:
100
mV
2. Connect the 5251 output to the Oscilloscope input. Terminate
the 5251output at the Oscilloscope input with the, 50
Ω
, 20 dB
Feed through termination
3. Configure the 5251as follows:
Frequency:
56.0000001
MHz
Output:
On
Adjustment:
1. Adjust CAL:SETUP44 to get the signal of 6 divisions on the
screen.
Setup 45
80 MHz Amplitude, Amplifier Out
Equipment: 50
Ω
, 20 dB Feed through termination, Oscilloscope
Preparation:
1. Configure the Oscilloscope as follows:
Input
Impedance:
50
Ω
Range:
100
mV
2. Connect the 5251 output to the Oscilloscope input. Terminate
the 5251output at the Oscilloscope input with the, 50
Ω
, 20 dB
Feed through termination
3. Configure the 5251as follows:
Frequency:
80
MHz
Output:
On
Adjustment:
1. Adjust CAL:SETUP45 to get the signal of 6 divisions on the
screen.
Setup 46
100 MHz Amplitude, Amplifier Out
Equipment: 50
Ω
, 20 dB Feed through termination, Oscilloscope
Preparation:
1. Configure the Oscilloscope as follows:
Input
Impedance:
50
Ω
Range:
100
mV
2. Connect the 5251 output to the Oscilloscope input. Terminate
the 5251output at the Oscilloscope input with the, 50
Ω
, 20 dB
Feed through termination
3. Configure the 5251as follows:
Frequency:
100
MHz
Output:
On
Adjustment:
1. Adjust CAL:SETUP46 to get the signal of 6 divisions on the
screen.
Содержание 5251
Страница 17: ...List of Figures continued xiv...
Страница 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Страница 37: ...TE5251 User Manual 2 2...
Страница 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Страница 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Страница 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Страница 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Страница 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Страница 290: ...TE5251 User Manual 2...