TE5251
User Manual
4-14
Instrument
Control
Commands
This group is used to control output channels and their respective
state, amplitude and offset settings, as well as the waveform mode.
You can also set the phase offset between channels and select
filters to re-structure the shape of your waveform. Multiple
instruments can be synchronized with these commands, as well.
The output frequency and the reference source are also selected
using commands from this group. Factory defaults after *RST are
shown in the Default column. Parameter range and low and high
limits are listed, where applicable.
Table
4-2, Instrument Control Commands Summary
Keyword
Parameter Range
Default
:OUTPut
:LOAD
50 to 1e6
50
[:STATe]
OFF | ON | 0 | 1
0
:SYNC
[:STATe]
OFF | ON | 0 | 1
0
:POSition
0 to 1999999
0
:FILTer
[:LPASs]
NONE | 25M | 50M | 60M | 120M
NONE
[:SOURce]
:ROSCillator
:SOURce
INTernal | EXTernal
INT
:FREQuency
[:CW]
10e-3 to 100e6 | MINimum | MAXimum
1e6
:RASTer
1.5 to 250e6 | MINimum | MAXimum
1e7
:SOURce
INTernal | EXTernal
INT
:VOLTage
[:LEVel]
[:AMPLitude]
100e-3 to 10 | MINimum | MAXimum
5
:OFFSet
-4.950 to 4.950
0
:FUNCtion
:MODE
FIXed | USER | SEQuence | MODulation | COUNter |
PULSe | HALFcycle
FIX
Содержание 5251
Страница 17: ...List of Figures continued xiv...
Страница 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Страница 37: ...TE5251 User Manual 2 2...
Страница 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Страница 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Страница 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Страница 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Страница 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Страница 290: ...TE5251 User Manual 2...