3. Configure the 5251as follows:
Frequency:
1
kHz
Output:
On
Amplitude:
1.1
V
Adjustment:
1. Adjust CAL:SETUP19 for DMM reading of 389 mV
±
3 mV
Setup 20
1 V Amplitude – Arbitrary, Amp-Out
Equipment: DMM, BNC to BNC cable, 50
Ω
Feed through termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
ACV
Range:
1
V
2. Connect the 5251 output to the DMM input. Terminate the
5251output at the DMM input with the, 50
Ω
Feed through
termination
3. Configure the 5251as follows:
Frequency:
1
kHz
Output:
On
Amplitude:
1
V
Adjustment:
1. Adjust CAL:SETUP20 for DMM reading of 353.5 mV
±
3 mV
Amplitude
Adjustments-
Modulation
The amplitude adjustments assure that the AC levels are within the
specified range. Use this procedure if you suspect that the amplitude
accuracy is an issue.
Setup 21
9 V Amplitude - Modulation
Equipment: DMM, BNC to BNC cable, 50
Ω
Feed through termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
ACV
Range:
10
V
2. Connect the 5251 output to the DMM input. Terminate the
5251output at the DMM input with the, 50
Ω
Feed through
termination
3. Configure the 5251as follows:
Frequency:
1
kHz
Output:
On
Содержание 5251
Страница 17: ...List of Figures continued xiv...
Страница 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Страница 37: ...TE5251 User Manual 2 2...
Страница 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Страница 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Страница 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Страница 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Страница 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Страница 290: ...TE5251 User Manual 2...