5
Performance Checks
Test Procedures
5-15
6. Configure the 5251, channel 1 only, as follows:
SCLK: 200
MS/s
Waveform: Arbitrary
Run Mode:
Triggered
Trigger Level
0 V
Trigger Delay:
On
Delay:
As required for the test
Amplitude: 5
V
Trigger Source: External
Output: On
Test Procedure:
1. Perform trigger delay tests using Tables 16
Table
Chapter 5-16, Trigger Delay Tests
5251 Delay Setting
Error Limits
Counter Reading Pass
Fail
1
μ
s 1
μ
s
±
150 ns
1 ms
1 ms
±
50 ns
1 s
1 s
±
50 ms
Re-trigger
Characteristics
Equipment: Counter, ArbConnection
Preparation:
1. Configure the counter as follows:
Function:
Pulse Width Measurement
Ch A Slope:
Negative
2. Connect the counter channel A to the 5251 output
3. Using ArbConnection prepare and download the following
waveform:
Wavelength: 100
points
Waveform:
Pulse, Delay = 0.1, Rise/Fall = 0, High Time
= 99.99
4. Configure the 5251 as follows:
SCLK 200
MS/s
Waveform: Arbitrary
Amplitude: 5
V
Run Mode:
Triggered
Trigger Level
0 V
Re-trigger: On
Re-trigger Delay: As required by the test
Trigger Source: Bus
Output: On
Test Procedure:
1. Manually trigger the instrument
2. Perform trigger delay tests using Tables 17
Содержание 5251
Страница 17: ...List of Figures continued xiv...
Страница 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Страница 37: ...TE5251 User Manual 2 2...
Страница 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Страница 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Страница 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Страница 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Страница 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Страница 290: ...TE5251 User Manual 2...