4
Remote Programming Reference
Instrument Control Commands
4-15
OUTPut:LOAD<load>
Description
This command will specify the load impedance that will be applied to the 5251 output.
Parameters
Name
Type
Default
Description
<load>
Numeric
(integer only)
50
Will specify the load impedance that will be applied to
the 5251 outputs in units of
Ω
. The default setting is
50
Ω
. The range of load impedance is 50
Ω
to 1
M
Ω
..Accurate setting of the load impedance is crucial
for correct display readout of the amplitude level on
the load.
OUTPut{OFF|ON|0|1}(?)
Description
This command will turn the 5251 output on and off. Note that for safety, the outputs always default to off,
even if the last instrument setting before power down was on
Parameters
Range
Type
Default
Description
0-1
Discrete
0
Sets the output on and off
Response
The 5251 will return 1 if the output is on, or 0 if the output is off.
OUTPut:SYNC{OFF|ON|0|1}(?)
Description
This command will turn the 5251 SYNC output on and off. Note that for safety, the SYNC output always
defaults to off, even if the last instrument setting before power down was on
Parameters
Range
Type
Default
Description
0-1
Discrete
0
Will set the SYNC output on and off
Response
The 5251 will return 1 if the SYNC output is on, or 0 if the SYNC output is off.
OUTPut:SYNC:POSition<position>(?)
Содержание 5251
Страница 17: ...List of Figures continued xiv...
Страница 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Страница 37: ...TE5251 User Manual 2 2...
Страница 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Страница 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Страница 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Страница 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Страница 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Страница 290: ...TE5251 User Manual 2...