TE5251
User Manual
6-20
3. Set oscilloscope input impedance to 50
Ω
4. Set oscilloscope vertical sensitivity to 0.1 V
Adjustment:
1. Adjust vertical trace to 6 divisions
2. Adjust C25 for best pulse response (4 ns type, 5 % aberrations)
Flatness
Adjustments
The flatness adjustments assure that the flatness of the amplifier is
within the specified range. Use this procedure if you suspect that the
flatness is an issue.
Setup 29
1 MHz Amplitude
Equipment: 50
Ω
, 20 dB Feed through termination, Oscilloscope
Preparation:
1. Configure the Oscilloscope as follows:
Input
Impedance:
50
Ω
Range:
100
mV
2. Connect the 5251 output to the Oscilloscope input. Terminate
the 5251output at the Oscilloscope input with the, 50
Ω
, 20 dB
Feed through termination
3. Configure the 5251as follows:
Frequency:
1
MHz
Output:
On
Adjustment:
1. Adjust the Fine Amplitude of the Oscilloscope to get the signal
of 6 divisions on the screen.
Setup 30
10 MHz Amplitude
Equipment: 50
Ω
, 20 dB Feed through termination, Oscilloscope
Preparation:
1. Configure the Oscilloscope as follows:
Input Impedance: 50 ohms
Range:
100
mV
2. Connect the 5251 output to the Oscilloscope input. Terminate
the 5251output at the Oscilloscope input with the, 50
Ω
, 20 dB
Feed through termination
3. Configure the 5251as follows:
Frequency:
10
MHz
Output:
On
Adjustment:
1. Adjust CAL:SETUP30 to get the signal of 6 divisions on the
screen.
Содержание 5251
Страница 17: ...List of Figures continued xiv...
Страница 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Страница 37: ...TE5251 User Manual 2 2...
Страница 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Страница 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Страница 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Страница 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Страница 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Страница 290: ...TE5251 User Manual 2...