TE5251
User Manual
5-16
Table
Chapter 5-17, Re-Trigger Delay Tests
5251 Re-trigger Setting
Error Limits Counter
Reading
Pass
Fail
1
μ
s 1
μ
s
±
150 ns
1 ms
1 ms
±
50 ns
1 s
1 s
±
50 ms
Trigger Slope
Equipment: Oscilloscope, function generator
Preparation:
1. Configure the Oscilloscope follows:
Termination:
50
Ω
, 20 dB feedthrough attenuator at the
oscilloscope input
Setup:
As required for the test
Trigger Source: External
2. Connect 5251 output to the oscilloscope input
3. Configure the function generator as follows:
Frequency 10
kHz
Run Mode:
Continue
Waveform:
TTL Output
4. Connect the function generator TTL output to the 5251 TRIG
IN connector
5. Connect the function generator main output to the 2
nd
channel
of the oscilloscope
6. Configure the 5251 as follows:
Frequency: 1
MHz
Waveform: Sine
wave
Run Mode:
Triggered
Output: On
Test Procedure:
1. Toggle 5251 trigger slope from positive to negative visa versa
2. Verify on the oscilloscope that the 5251 transitions are
synchronized with the slope of the trigger
Test Results
Pass
Fail
Trigger Level
Equipment: Oscilloscope, function generator
Preparation:
1. Configure the Oscilloscope as follows:
Termination:
50
Ω
, 20 dB feedthrough attenuator at the
oscilloscope input
Setup:
As required for the test
2. Connect 5251 output to the oscilloscope input
3. Configure the function generator as follows:
Содержание 5251
Страница 17: ...List of Figures continued xiv...
Страница 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Страница 37: ...TE5251 User Manual 2 2...
Страница 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Страница 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Страница 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Страница 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Страница 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Страница 290: ...TE5251 User Manual 2...