Performance Checks
Test Procedures
5-17
Frequency 10
kHz
Run Mode:
Continuous
Waveform: Squarewave.
Amplitude:
1 V
4. Connect the function generator output to the 5251 TRIG IN
connector
5. Configure the 5251 as follows:
Frequency: 1
MHz
Waveform: Sine
wave
Run Mode:
Triggered
Trigger level:
0 V
Ch1 Output:
On
Test Procedure:
1. Verify that the 5251 outputs triggered waveforms spaced at
0.1 ms
2. Modify the function generator offset to +2 V and change the
5251 trigger level to +4 V. Verify that the 5251 triggered
waveforms are spaced 0.1 ms apart
3. Modify the function generator offset to -2 V and change the
5251 trigger level to -4 V. Verify that the 5251 triggered
waveforms are spaced 0.1 ms apart
Test Results
Pass
Fail
Sequence
operation
This tests the operation of the sequence generators. This also checks
the various sequence advance options.
Automatic Advance
Equipment: Counter
Preparation:
1. Configure the Counter as follows:
Function: TOTB
Measurement
2. Connect the counter channel B to the 5251 output
3. Configure the 5251 as follows:
SCLK 200
MS/s
Waveform: Sequence
Run Mode:
Trigger
Amplitude: 2
V
Output: On
4. Using ArbConnection prepare and download the following
waveform:
Segments:
1 to 5
Содержание 5251
Страница 17: ...List of Figures continued xiv...
Страница 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Страница 37: ...TE5251 User Manual 2 2...
Страница 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Страница 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Страница 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Страница 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Страница 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Страница 290: ...TE5251 User Manual 2...