TA I
NSTRUMENTS
TMA 2940
2–25
Air Cool Line
1. Locate the COOLING GAS fitting, a 6.2-
mm (1/4-inch) compression fitting on the
left side of the TMA cabinet back, marked
with a maximum pressure warning label
(840 kPa [120 psi]), as seen in the figure
below.
Figure 2.15
TMA COOLING
GAS Fitting
2. Make sure your compressed lab air source is
dry, filtered, and regulated to between 175
kPa (25 psi) and 840 kPa (120 psi).
3. Connect a compressed air line to the COOL-
ING GAS fitting.