Using Your Options
TA I
NSTRUMENTS
TMA 2940
4–22
To calculate the vial constant from a scan use
the following equation:
b
ref
V
K =
(S B)
where:
K
= vial constant (cm
3
/
µ
m)
S
= slope (
µ
m/
o
C)
b
ref
= cubic expansion coefficient of
standard material
m
3
m
3 o
C
b
ref
= 3
a
ref
Where
a
ref
is the coefficient of
linear expansion of standard
material, (m/m
o
C), measured
over the same temperature range.
V
= sample volume (cm
3
). This volume
may be obtained by dividing the
sample mass in grams by its density
in g/cm
3
at room temperature.
B
= slope of baseline (
µ
m/
o
C).
The dilatometer accessory kit contains:
Dilatometer probe assembly
Dilatometer sample vial
Filling medium
Aluminum calibration standards.
( )
Содержание TMA 2940
Страница 18: ...TA INSTRUMENTS TMA 2940 XVIII...
Страница 20: ...Introduction 1 2 TA INSTRUMENTS TMA 2940...
Страница 34: ...Introduction 1 16 TA INSTRUMENTS TMA 2940...
Страница 36: ...Installation TA INSTRUMENTS TMA 2940 2 2...
Страница 62: ...Installation TA INSTRUMENTS TMA 2940 2 28 Figure 2 17 Isolation Transformer Cable Connections...
Страница 66: ...Installation TA INSTRUMENTS TMA 2940 2 32...
Страница 138: ...Using Your Options TA INSTRUMENTS TMA 2940 4 48...
Страница 140: ...5 2 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 150: ...5 12 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 152: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 2...
Страница 168: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 18...