Displacement
Calibration
Calibrate the dilatometer for cubic displacement
using the following procedure:
1. Take three aluminum balls from the dilatom-
eter accessory kit to use as a calibration
sample.
2. Weigh the three aluminum balls and prepare
the sample and vial as instructed on the
following pages. Make sure that the alumi-
num balls do not touch the walls of the vial,
and that they do not touch ech other.
3. Obtain a thermal scan under the same
conditions of the baseline scan (
e.g.
same
load and heating rate).
4. Calculate the vial constant from the scan by
using the equation found on page 4-22.