Maintenance and Diagnostics
TA I
NSTRUMENTS
TMA 2940
6–8
Replacements
Occasionally you may need to replace a broken
or worn out part of the TMA. Any replacements
needed, other than those discussed in this
manual, must be supplied and installed by
qualified TA Instruments service personnel. Call
(302) 427-4050 for service.
Installing a
New Thermocouple
You may find that the need arises for a new
thermocouple (PN 944017.901) as a result of
normal wear and tear, accidental breakage,
contamination of the thermocouple,
etc
.
Before removing the thermocouple, you will
need to take off the TMA balance enclosure, the
stage shield, and the spring clip holding the
thermocouple in place. Then carefully thread the
thermocouple down through the opening in the
platform and unplug it.
To replace the sample thermocouple, follow the
directions found in Chapter 2, Installation.
Содержание TMA 2940
Страница 18: ...TA INSTRUMENTS TMA 2940 XVIII...
Страница 20: ...Introduction 1 2 TA INSTRUMENTS TMA 2940...
Страница 34: ...Introduction 1 16 TA INSTRUMENTS TMA 2940...
Страница 36: ...Installation TA INSTRUMENTS TMA 2940 2 2...
Страница 62: ...Installation TA INSTRUMENTS TMA 2940 2 28 Figure 2 17 Isolation Transformer Cable Connections...
Страница 66: ...Installation TA INSTRUMENTS TMA 2940 2 32...
Страница 138: ...Using Your Options TA INSTRUMENTS TMA 2940 4 48...
Страница 140: ...5 2 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 150: ...5 12 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 152: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 2...
Страница 168: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 18...