Drying out the instrument may be needed, if it has
been exposed to humid conditions. Certain ceramic
materials used in this equipment may absorb
moisure, causing leakage currents to exceed those
specified in the applicable standards (see page xiii)
until the moisture is eliminated. It is important to
be certain that the instrument ground is ad-
equately connected to the facilities ground for safe
operation.
Run the following method to dry out the instrument
(refer to “Running Experiments” for further infor-
mation).
1
Ramp at 10°C/min to 400°C
2
Isothermal for 30 min.
Installing the
Thermocouple
Your TMA 2940 will arrive without a sample
thermocouple installed. With the balance
enclosure removed and the furnace raised,
follow these steps:
1. Insert the two-pin, color-coded thermo-
couple connector into the fitting as shown in
Figure 2.5.
Figure 2.5
Plugging in
the Thermocouple
Installing the Instrument
NOTE:
Содержание TMA 2940
Страница 18: ...TA INSTRUMENTS TMA 2940 XVIII...
Страница 20: ...Introduction 1 2 TA INSTRUMENTS TMA 2940...
Страница 34: ...Introduction 1 16 TA INSTRUMENTS TMA 2940...
Страница 36: ...Installation TA INSTRUMENTS TMA 2940 2 2...
Страница 62: ...Installation TA INSTRUMENTS TMA 2940 2 28 Figure 2 17 Isolation Transformer Cable Connections...
Страница 66: ...Installation TA INSTRUMENTS TMA 2940 2 32...
Страница 138: ...Using Your Options TA INSTRUMENTS TMA 2940 4 48...
Страница 140: ...5 2 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 150: ...5 12 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 152: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 2...
Страница 168: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 18...