![TA Instruments TMA 2940 Скачать руководство пользователя страница 133](http://html1.mh-extra.com/html/ta-instruments/tma-2940/tma-2940_operators-manual_3130625133.webp)
TA I
NSTRUMENTS
TMA 2940
4–43
Thermosetting material
Soak the plates and cages in a suitable solvent to
separate and clean them.
or
Heat the parts to red heat in the oxidizing
portion of a burner flame. The organic material
should oxidize completely.
or
Heat the plates in a furnace to approximately
1000°C.
Calculations
Correction Calculations for Thermal
Expansion of Parallel Plates
1. Calculate the apparent distance between the
plates (h*) at a given temperature (or time)
by subtracting the decrease in distance
between the plates (in mm) from the original
sample height.
Equation:
h* = sample thickness – dimension change
(mm)
(mm)
2. Correct the distance between the plates for
thermal expansion.
Each plate is nominally 5.08 mm (0.2 inch)
stainless steel. Therefore, the total dimen-
sion for both plates is 10.2 mm (0.4 in.).
Parallel Plate Rheometer Accessory
Содержание TMA 2940
Страница 18: ...TA INSTRUMENTS TMA 2940 XVIII...
Страница 20: ...Introduction 1 2 TA INSTRUMENTS TMA 2940...
Страница 34: ...Introduction 1 16 TA INSTRUMENTS TMA 2940...
Страница 36: ...Installation TA INSTRUMENTS TMA 2940 2 2...
Страница 62: ...Installation TA INSTRUMENTS TMA 2940 2 28 Figure 2 17 Isolation Transformer Cable Connections...
Страница 66: ...Installation TA INSTRUMENTS TMA 2940 2 32...
Страница 138: ...Using Your Options TA INSTRUMENTS TMA 2940 4 48...
Страница 140: ...5 2 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 150: ...5 12 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 152: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 2...
Страница 168: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 18...