![TA Instruments TMA 2940 Скачать руководство пользователя страница 136](http://html1.mh-extra.com/html/ta-instruments/tma-2940/tma-2940_operators-manual_3130625136.webp)
Using Your Options
TA I
NSTRUMENTS
TMA 2940
4–46
η
=
Calculations for Viscosity and Wall
Shear Rate Data
To calculate sample viscosity (
η
) and wall shear
rate (
γ
), the experimental data is first scaled and
corrected for expansion of the parallel plates (as
previously shown on page 4-43). Then this
corrected data is applied to equations developed
by Dienes and Klemm (J. Appl. Phys., 17,458,
1946) and Cessna and Jabloner (J. Elast. Plast.,
6, 103, 1974).
Equation:
4F
∆
t
x
x
10
6
mm
2
/m
2
3
π
r
4
∆
(1/h
2
)
3rh
∆
(1/h
2
)
2
∆
t
where:
F
=
applied force (N)
r
=
parallel plate radius (mm)
h
=
sample height (mm)
t
=
time (sec)
η
=
viscosity (Pa-sec)
γ
=
shear rate (1/sec)
γ
=
x
•
•
Содержание TMA 2940
Страница 18: ...TA INSTRUMENTS TMA 2940 XVIII...
Страница 20: ...Introduction 1 2 TA INSTRUMENTS TMA 2940...
Страница 34: ...Introduction 1 16 TA INSTRUMENTS TMA 2940...
Страница 36: ...Installation TA INSTRUMENTS TMA 2940 2 2...
Страница 62: ...Installation TA INSTRUMENTS TMA 2940 2 28 Figure 2 17 Isolation Transformer Cable Connections...
Страница 66: ...Installation TA INSTRUMENTS TMA 2940 2 32...
Страница 138: ...Using Your Options TA INSTRUMENTS TMA 2940 4 48...
Страница 140: ...5 2 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 150: ...5 12 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 152: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 2...
Страница 168: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 18...