Using Your Options
TA I
NSTRUMENTS
TMA 2940
4–12
Sample
Preparation
The method used to prepare and load your
sample varies according to the sample type.
There are two methods available for the film/
fiber accessory: aluminum balls or stainless
steel clamps.
Using the
Aluminum Balls
The aluminum balls are generally used with
fiber samples as follows:
1. Measure and cut the fiber sample slightly
longer than desired length, typically 12 to 25
mm (0.5 to 1.0 inch).
2. Place two cleaved aluminum balls 12 to 25
mm (0.5 to 1.0 inch) apart.
3. Lay the fiber across the balls and press the
fiber into the cleavage of each ball (see
Figure 4.3).
Figure 4.3
Fiber Sample
Preparation
Sample Length
12 to 25 mm
(0.5 to 1.0 inch)
Содержание TMA 2940
Страница 18: ...TA INSTRUMENTS TMA 2940 XVIII...
Страница 20: ...Introduction 1 2 TA INSTRUMENTS TMA 2940...
Страница 34: ...Introduction 1 16 TA INSTRUMENTS TMA 2940...
Страница 36: ...Installation TA INSTRUMENTS TMA 2940 2 2...
Страница 62: ...Installation TA INSTRUMENTS TMA 2940 2 28 Figure 2 17 Isolation Transformer Cable Connections...
Страница 66: ...Installation TA INSTRUMENTS TMA 2940 2 32...
Страница 138: ...Using Your Options TA INSTRUMENTS TMA 2940 4 48...
Страница 140: ...5 2 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 150: ...5 12 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 152: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 2...
Страница 168: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 18...