Running Experiments
TA I
NSTRUMENTS
TMA 2940
3–14
5. Place the sample on the stage under the
probe tip (see Figure 3.1).
It is recommended that you place a quartz wafer
or a piece of thin aluminum foil between the stage
and any thermoplastic samples to prevent damage
to the stage.
Figure 3.1
Loading a Sample
6. Adjust the sample thermocouple, if needed,
as directed on the next page.
7. Program the force on the controller ,or add
weight to the weight tray, if desired. Be sure
to enter any static weight added to the
weight tray on the instrument control pro-
gram; see the online documentation for
details.
NOTE: