TA I
NSTRUMENTS
TMA 2940
4–9
6. Attach the thermocouple to the stage as
follows:
a. Position the tip of the thermocouple
approximately one half the length of the
samples you will be measuring. The tip
should be angled toward the sample
location.
After installing the thermocouple, tip placement
may interfere with the zero length operation.
Temporarily move the thermocouple when per-
forming this operation.
b. Hold the thermocouple against the stage
assembly and put on the spring clip to
hold the thermocouple in place.
7. Replace the stage shield on the base of the
stage.
8. Move the furnace back into position and
lower it. See Figures 4.4 and 4.7 for the
correct probe setup with fiber and film
samples.
Calculating
the Offset
When you use the film/fiber accessory, analysis
must compensate for the thickness of the stage
platform and the upper portion of the probe.
Perform the Auto zero offset procedure to
compensate for the thickness. (See the
Thermal
Solutions/Advantage User Reference Guide
for further instructions on performing this
operation.)
Film/Fiber Accessory
NOTE:
Содержание TMA 2940
Страница 18: ...TA INSTRUMENTS TMA 2940 XVIII...
Страница 20: ...Introduction 1 2 TA INSTRUMENTS TMA 2940...
Страница 34: ...Introduction 1 16 TA INSTRUMENTS TMA 2940...
Страница 36: ...Installation TA INSTRUMENTS TMA 2940 2 2...
Страница 62: ...Installation TA INSTRUMENTS TMA 2940 2 28 Figure 2 17 Isolation Transformer Cable Connections...
Страница 66: ...Installation TA INSTRUMENTS TMA 2940 2 32...
Страница 138: ...Using Your Options TA INSTRUMENTS TMA 2940 4 48...
Страница 140: ...5 2 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 150: ...5 12 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 152: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 2...
Страница 168: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 18...