TA I
NSTRUMENTS
TMA 2940
4–13
4. Crimp the two balls closed. The balls should
not shift or slip when tested by hand.
5. Cut off any excess fiber extending beyond
the aluminum balls.
6. Place one ball over the slot in the stage with
the fiber hanging down through the slot and
allow the other ball to hang freely.
7. Position the second ball below the slot in the
film/fiber probe so that the fiber is sus-
pended between the two balls as shown in
Figure 4.4.
8. Measure the sample length.
9. Position the thermocouple close to, but not
touching the sample.
Figure 4.4
Probe Setup
Using Fiber Samples
Film/Fiber Accessory