Because of its conductivity, helium is the
purge gas recommended when performing
TMA measurements at temperatures below
ambient. Helium reduces the time required
to cool the sample, provides a more stable
sample environment, and reduces cold
drafts on the system mechanism.
•
Use the following helium purge rates for
subambient operation:
— 100 mL/min at -100
o
C and above
— 200 mL/min at -150
o
C and above.
•
If a denser purge gas than helium is used for
subambient experiments, it is recommended
that the silicon washer found in the
accessory kit be attached to the probe.
Raise the probe to the fully open position,
then place the washer so it is just below the
opening in the stage. This will minimize
cold drafts into the lower balance enclosure.
•
If slow heating rates or isothermal holds are
desired at subambient temperatures, it is
recommended that the aluminum mass
found in the accessory kit be inserted into
the furnace reservoir. This adds a more
stable cooling mass for the furnace to work
against.
Do not program furnace temperatures
greater than 600
o
C when using the aluminum
thermal mass.
t
t
t
t
t
CAUTION:
Содержание TMA 2940
Страница 18: ...TA INSTRUMENTS TMA 2940 XVIII...
Страница 20: ...Introduction 1 2 TA INSTRUMENTS TMA 2940...
Страница 34: ...Introduction 1 16 TA INSTRUMENTS TMA 2940...
Страница 36: ...Installation TA INSTRUMENTS TMA 2940 2 2...
Страница 62: ...Installation TA INSTRUMENTS TMA 2940 2 28 Figure 2 17 Isolation Transformer Cable Connections...
Страница 66: ...Installation TA INSTRUMENTS TMA 2940 2 32...
Страница 138: ...Using Your Options TA INSTRUMENTS TMA 2940 4 48...
Страница 140: ...5 2 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 150: ...5 12 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 152: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 2...
Страница 168: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 18...