Using Your Options
TA I
NSTRUMENTS
TMA 2940
4–28
Figure 4.10
Probe Setup Using
Dilatometer Samples
14. Press
ZERO LENGTH
on the keypad to
provide a zero reference point.
15. Adjust the thermocouple position so that it
is in close proximity to the vial.
To avoid false thermocouple readings, make sure
that the thermocouple tip does not touch the
sample vial.
16. Move the furnace into position and lower it.
17. Program the force (typically 0.01 N or less)
using
Thermal Solutions/Advantage
.
18. Check that the sample size is equal to zero
before beginning your experiment.
NOTE: