Running Experiments
TA I
NSTRUMENTS
TMA 2940
3–22
Subambient
Experiments
The TMA can be used to run experiments on
cooled samples with a cooling source such as
liquid nitrogen. In order to perform subambient
experiments the furnace must work against the
coolant in the reservoir. Ideally, you should use
the least amount of coolant necessary to keep the
furnace drawing power over the entire sub-
ambient range.
If the “Return to temperature range” method-end
condition is activated in the instrument control program,
press the STOP key to deactivate it before trying to cool
the TMA.
Performing a
Cooling Experiment
To run a subambient experiment perform the
same operations as described in "Running a
TMA Experiment" on pages 3-10 to 3-20. Once
the sample is loaded and the furnace is lowered,
conduct the following operations:
1. Turn on the purge gas at 100 mL/min and
allow it to flow for 3 to 5 minutes to ensure
a moisture-free environment.
2. Pour coolant carefully into the top of the
furnace, refilling it as needed in order to
cool the sample below the desired tempera-
ture.
NOTE:
Содержание TMA 2940
Страница 18: ...TA INSTRUMENTS TMA 2940 XVIII...
Страница 20: ...Introduction 1 2 TA INSTRUMENTS TMA 2940...
Страница 34: ...Introduction 1 16 TA INSTRUMENTS TMA 2940...
Страница 36: ...Installation TA INSTRUMENTS TMA 2940 2 2...
Страница 62: ...Installation TA INSTRUMENTS TMA 2940 2 28 Figure 2 17 Isolation Transformer Cable Connections...
Страница 66: ...Installation TA INSTRUMENTS TMA 2940 2 32...
Страница 138: ...Using Your Options TA INSTRUMENTS TMA 2940 4 48...
Страница 140: ...5 2 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 150: ...5 12 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 152: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 2...
Страница 168: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 18...