3–21
To clean off polymeric material,
carefully apply heat to the stage and
probe area, removing the sample as it
softens. Take care not to overheat
mineral- or glass-filled polymers as this
will cause the filler to stick to the stage
or probe.
To remove metallic materials (
e.g.
,
indium), immerse the probe and stage in
an acid solution.
Use extreme caution when working with
acid solutions.
Heat is not recommended with metallic
materials, as heat causes the metal to
diffuse into the quartz and will change
the properties of the stage and probe.
Running a TMA Experiment
t
CAUTION: