TA I
NSTRUMENTS
TMA 2940
4–11
4. Press the
Measure Length
key on the
keypad. Record this value as the
Displayed
Size.
5. Calculate the new offset using the following
equation:
Offset
new
= (Actual Size - Displayed Size) +
Offset
current
6. Enter the new Auto Measure Offset using
the
Thermal Solutions/Advantage
Instru-
ment Control functions.
7. Remove the Auto Measure Gauge and
proceed with your experiment.
Calibration
You should perform probe, temperature, and cell
constant calibration for the film/fiber accessory
as you would normally calibrate any of the
standard probes. Metal wires crimped in the
aluminum balls should be used for temperature
calibration. High quality aluminum foil
clamped in the steel clamps should be used for
cell constant calibration. See the
Thermal
Solutions/Advantage User Reference Guide
for details.
Film/Fiber Accessory
Содержание TMA 2940
Страница 18: ...TA INSTRUMENTS TMA 2940 XVIII...
Страница 20: ...Introduction 1 2 TA INSTRUMENTS TMA 2940...
Страница 34: ...Introduction 1 16 TA INSTRUMENTS TMA 2940...
Страница 36: ...Installation TA INSTRUMENTS TMA 2940 2 2...
Страница 62: ...Installation TA INSTRUMENTS TMA 2940 2 28 Figure 2 17 Isolation Transformer Cable Connections...
Страница 66: ...Installation TA INSTRUMENTS TMA 2940 2 32...
Страница 138: ...Using Your Options TA INSTRUMENTS TMA 2940 4 48...
Страница 140: ...5 2 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 150: ...5 12 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 152: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 2...
Страница 168: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 18...