Status Codes
Status Codes
Status codes are character strings that are
continuously displayed at the top left of the
TMA instrument display. These codes tell you
what segment is currently being performed by
the instrument.
Table 5.1
Method Status Codes
Code
Meaning
Air Cool
The air cool solenoid is
opened to allow cooling
air flow to the furnace.
Balance
Force is being applied
to balance the system at
the center of the displace-
ment range. This is used
for force calibration.
Calib
The instrument is per-
forming LVDT calib-
ration or the instrument is
in a calibration mode. No
method is running.
Cold
The instrument heater
cannot supply heat fast
enough to keep up with
the thermal program.
This may be caused by a
large ballistic jump in the
program, a faulty heater, a
faulty control thermo-
couple signal.
(table continued)
Содержание TMA 2940
Страница 18: ...TA INSTRUMENTS TMA 2940 XVIII...
Страница 20: ...Introduction 1 2 TA INSTRUMENTS TMA 2940...
Страница 34: ...Introduction 1 16 TA INSTRUMENTS TMA 2940...
Страница 36: ...Installation TA INSTRUMENTS TMA 2940 2 2...
Страница 62: ...Installation TA INSTRUMENTS TMA 2940 2 28 Figure 2 17 Isolation Transformer Cable Connections...
Страница 66: ...Installation TA INSTRUMENTS TMA 2940 2 32...
Страница 138: ...Using Your Options TA INSTRUMENTS TMA 2940 4 48...
Страница 140: ...5 2 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 150: ...5 12 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 152: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 2...
Страница 168: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 18...