Starting the 2940
Allow the TMA to warm up for at least 30 minutes
before performing an experiment.
1. Check all connections between the TMA
and the controller. Make sure each compo-
nent is plugged into the correct connector, as
described in this chapter.
2. Turn the instrument
POWER
switch to the
ON ( ) position. The instrument will run an
internal confidence test each time you turn
on the power.
3. Watch the instrument display during the
confidence test for any messages that may
be indicated. Confidence test definitions are
given in Chapter 6.
After the confidence test has finished, the screen
will briefly display the system status, indicating
the amount of data storage memory available
and the GBIP address. Next follows the copy-
right display, and then the standby display,
shown in the figure on the next page.
4. Bring the instrument online with the TA
controller.
NOTE:
Starting the 2940
Содержание TMA 2940
Страница 18: ...TA INSTRUMENTS TMA 2940 XVIII...
Страница 20: ...Introduction 1 2 TA INSTRUMENTS TMA 2940...
Страница 34: ...Introduction 1 16 TA INSTRUMENTS TMA 2940...
Страница 36: ...Installation TA INSTRUMENTS TMA 2940 2 2...
Страница 62: ...Installation TA INSTRUMENTS TMA 2940 2 28 Figure 2 17 Isolation Transformer Cable Connections...
Страница 66: ...Installation TA INSTRUMENTS TMA 2940 2 32...
Страница 138: ...Using Your Options TA INSTRUMENTS TMA 2940 4 48...
Страница 140: ...5 2 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 150: ...5 12 TA INSTRUMENTS TMA 2940 Technical Reference...
Страница 152: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 2...
Страница 168: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 18...