Mounting Location
Installation
Mounting Location
The ambient temperature for operating the T-Comfort 930 DSL must be b5 and +40°C. The power
supply must be 230 V/50 Hz AC. A separate fuse for the power supply is recommended.
To maintain the prescribed ambient temperature, mount the T-Comfort 930 DSL in a well-ventilated location,
away from direct sources of heat.
Do not position the T-Comfort 930 DSL
■
in front of or above heat sources such as radiators,
■
in direct sunlight,
■
behind curtains,
■
in small, unventilated, damp rooms,
■
on or near inflammable materials,
■
or near high-frequency devices such as transmitters, X-ray or similar apparatus.
Use a separate 230 V power circuit and install overvoltage protection.