![SRS Labs QMS 100 Series Скачать руководство пользователя страница 99](http://html1.mh-extra.com/html/srs-labs/qms-100-series/qms-100-series_user-manual_1354225099.webp)
Capillary Design 1-11
Technical Reference
The chromatography industry uses a large variety of capillaries, from which we can select capillaries for
the QMS. The figure below shows the conductivity for several commonly available bore diameters.
1E-08
1E-07
1E-06
1E-05
1E-04
1E-03
1E-02
1
10
100
1000
length (cm)
C (liter/s)
2
4
5
6
7
Figure 6. Conductivity of five different capillaries as a function of length. The curves are labeled with the bore
diameter in thousandths of an inch. Both axis are logarithmic.
As an example, consider a capillary for atmospheric pressure and an capillary exit pressure of 1 mbar.
From the speed curve for the diaphragm pump, the throughput,
Q
, is 2.5
×
10
-2
mbar liter s
-1
(
Q
=
P S
)
and the required conductivity is 2.5
×
10
-5
liter s
-1
. The horizontal dotted line in figure 6 shows that for
each capillary diameter there is an appropriate length. The application dictates which bore diameter is
appropriate. If the system being measured was far away from the QMS, 3 meters of the 7/1000
capillary would be best. On the other hand, for something close 0.7 meters of 5/1000 would be less
expensive. The gas velocity through the capillary quickly reaches the speed of sound. Thereby, until the
capillary becomes
very
long, it is not an important contributor to the response time of the QMS.
Содержание QMS 100 Series
Страница 5: ...v SRS QMS Gas Analyzer...
Страница 79: ...Technical Reference Manual QMS 100 Series Gas Analyzer Version 3 2 1 2012...
Страница 102: ......
Страница 124: ......
Страница 194: ......
Страница 216: ...5 8 Accessing Internal Components SRS QMS Gas Analyzer Internal View Map of Internal Components...
Страница 238: ......